SAXS experiment

SAXS is a contrast method where the scattering signal is obtained from the difference in electron density between a molecule in solution () and bulk solvent (), scattering resulting from the buffer has to be subtracted from the measured scattering of the sample to isolate the sole contribution from the molecule (see the Basic SAXS data processing section).

Sample requirement

TL;DR: at least 50 µL of sample at a concentration of 1 – 10 mg/ml, along with 1 ml of matching buffer

Because of the nature of the SAXS method, sample (molecule + buffer) and buffer (buffer alone) must match. In order to achieve this, dialyzing the sample overnight in the buffer of choice is the best method.

Alternatively, yet not recommended, size-exclusion chromatography buffer or diafiltration flow-through could be used for approximate buffer blank. However, adequate buffer subtraction is not guaranteed and results may be of poor quality forbidding downstream processing of the data, or give erroneous information about the sample quality. Use these alternative methods at your own risk!

Also, if possible, avoid high concentrations of osmolytes (glycine, betaine, sucrose, glycerol etc) as they will decrease the scattering contrast with the sample. High salt concentrations above 1 M should also be avoided for the same reason.

The sample needs to be monodisperse. This can easily be estimated by gel filtration, non-denaturing gel electrophoresis or dynamic light scattering.

Appropriate concentration of sample for good scattering signal: 1 – 10 mg/ml (lower concentrations will not give meaningful data) for proteins. Nucleic acids scatter X-rays approximately 5 times more than proteins (because of their higher electron density), thus allowing lower concentrations. At least 50 µl of sample (more is better) is required for SAXS analysis and preparation of dilution series, and 1 ml of matching buffer should be sufficient. If detergents are required for the stability of the sample, avoid detergent concentrations above critical micelle concentration (CMC) (some exceptions may apply).

Once you know your sample can meet the SAXS requirements, you can reserve some time on the instrument by contacting me by email (normand.cyr@umontreal.ca). I need to know at least 8 hours in advance in order to start up the X-ray source.

Sample preparation

On the day of the SAXS experiment:

Setup

Setup the plate configuration on the computer:

Setup the run program:

Data acquisition

Once the plate and the program are set, the data acquisition can be started by pressing the start button.

Radial averaging of the scattering

X-ray scattering will be detected and a 2D image representing the scattering intensity over the detector plane will be recorded. Then, from the center of the beam, radial averaging of the intensity is performed at a specific q-value interval. The resulting average intensity (I) is finally plot against the q-value (q) to generate the scattering curve.